Agilent Technologies 87020 Stroller User Manual


 
6-69
Performance Tests Option 011
Performance Test Records - Option 011
Performance Test Records - Option 011
The following tables are to record the performance test records for each of the
Agilent 8702D Option 011 tests.
Test 1. Source Frequency Range and Accuracy
CW Frequency
(MHz)
Lower Limit
(MHz)
Measured Value
(MHz)
Upper Limit
(MHz)
Measurement
Uncertainty (MHz)
0.3 0.299 997
____________________
0.300 003 ± 0.000 000 360
5.0 4.999 950
____________________
5.000 050 ± 0.000 006
16.0 15.999 840
____________________
16.000 160 ± 0.000 019
31.0 30.999 690
____________________
31.000 310 ± 0.000 037
60.999 999 60.999 390
____________________
61.000 610 ± 0.000 073
121.0 120.998 790
____________________
121.001 210 ± 0.000 145
180.0 179.998 200
____________________
180.001 800 ± 0.000 216
310.0 309.995 900
____________________
310.003 100 ± 0.000 372
700.0 699.930 000
____________________
700.007 000 ± 0.000 840
1 300.0 1 299 987
____________________
1 300 013 ± 0.001 560
2 000.0 1 999 980
____________________
2 000 020 ± 0.002 400
3 000.0 2 999 970
____________________
3 000 030 ± 0.003 600
Option 006
4.0 3.999 960
____________________
4.000 040 ± 0.004 800
5.0 4.999 950
____________________
5.000 050 ± 0.006 000
6.0 5.999 940
____________________
6.000 060 ± 0.007 200