M400E Ozone Analyzer Operator’s Manual Advanced FEATURES of the M400E analyzer
7.1.5.4. Editing Sample Period and Report Period
The iDAS defines two principal time periods by which sample readings are taken and permanently recorded:
SAMPLE PERIOD: Determines how often iDAS temporarily records a sample reading of the parameter
in volatile memory. The
SAMPLE PERIOD is set to one minute by default and generally cannot be
accessed from the standard iDAS front panel menu, but is available via the instruments communication
ports by using APICOM or the analyzer’s standard serial data protocol.
SAMPLE PERIOD is only used
when the iDAS parameter’s sample mode is set for AVG, MIN or MAX.
REPORT PERIOD: Sets how often the sample readings stored in volatile memory are processed, (e.g.
average, minimum or maximum are calculated) and the results stored permanently in the instruments
Disk-on-Chip as well as transmitted via the analyzer’s communication ports. The
REPORT PERIOD may
be set from the front panel. If the INST sample mode is selected the instrument stores and reports an
instantaneous reading of the selected parameter at the end of the chosen report period.
To define the
REPORT PERIOD, follow the instruction shown in Section 7.1.5 then press: