357BRemote Operation Of The M400E M400E Ozone Analyzer Operator’s Manual
8.1.6. MACHINE ID
Each type of Teledyne Instruments’ analyzer is configured with a default ID code. The default ID code for all
M400E analyzers is
700. The ID number is only important if more than one analyzer is connected to the same
communications channel such as when several analyzers are on the same Ethernet LAN (See Section 8.4); in a
RS-2
32 multidrop chain (See Section 8.2.1) or operating over a RS-485 network (See Section 8.3). If two
analyzers of the sam
e model type are used on one channel, the
ID codes of one or both of the instruments
needs to be changed so
To edit the instrument’s ID code, press:
SETUP X.X COMMUNICATIONS MENU
ID COM1 COM2 EXIT
SETUP X.X MACHINE ID:400 ID
0 4 0 0 ENTR EXIT
Toggle these keys to
cycle through the
available character
set: 0-7
EXIT discards the new
setting
ENTR accepts the
new setting
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE=500.0 PPB O3= XXXX
<TST TST> CAL SETUP
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
The ID number is only important if more than one analyzer is connected to the same communications channel
(e.g., a multi-drop setup). Different models of Teledyne Instruments’ analyzers have different default ID
numbers, but if two analyzers of the same model type are used on one channel (for example, two M400E’s), the
ID of one instrument needs to be changed.
The ID can also be used for to identify any one of several analyzers attached to the same network but situated in
different physical locations.