Teledyne 400E Dollhouse User Manual


 
M400E Calibration Procedures M400E Ozone Analyzer Operator’s Manual
EXIT
returns
to the SETUP
Menu
SETUP X.X DURATION:15.0 MINUTES
<SET SET> EDIT EXIT
SETUP X.X
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP X.X DURATION:30.0 MINUTES
<SET SET> EDIT EXIT
Toggle keys to set
duration for each iteration
of the sequence:
Set in Decimal minutes
from 0.1 – 60.0
SETUP X.X CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP X.X CALIBRATE: OFF
ON ENTR EXIT
SETUP X.X C
A
LIBRATE: ON
<SET SET> EDIT
EXIT
Toggle key
Between Off and
ON
SETUP X.X
SEQ 2) ZERO
SPAN, 2:00:30
PREV NEXT MODE SET EXIT
Display show:
SEQ 2) ZERO–SPAN, 2:00:30
Sequence Delta Time
MODE Delta Days
CONTINUED FROM PREVIOUS PAGE
DELTA TIME
NOTE
If at any time an illegal entry is selected (Example: Delta Days > 367) the ENTR key will disappear from
the display.
166 04315 Rev. C1