8.4.3. CHANGING THE ANALYZER’S HOSTNAME
The HOSTNAME is the name by which the analyzer appears on your network. The default name for all
Teledyne Instruments M400E analyzers is
M400E. To change this name (particularly if you have more than one
M400E analyzer on your network), press.
Use these key to edit the HOSTNAME
SETUP X.X COMMUNICATIONS MENU
ID ADDR INET EXIT
SETUP X.X ENTER PASSWORD:818
81 8 ENTREXIT
SETUP X.X DHCP:ON
<SET SET> EDIT EXIT
SETUP X.X HOSTNAME: TMS 9000
<SET SET> EDIT EXIT
SETUP X.X INITIALIZING INET 0%
INITIALIZATION process proceeds
automatically
SETUP X.X COMMUNICATIONS MENU
ID ADDR INET EXIT
SETUP X.X INITIALIZATION SUCCEEDED SETUP X.X INITIALIZATION FAILED
Contact your
IT Network
Administrator
Continue pressing SET> until ...
SETUP X.X HOSTNAME: TMS 9000
<CH CH> INS DEL [?] ENTR EXIT
SETUP X.X HOSTNAME: TMS 9K–STACK 2
<CH CH> INS DEL [?] ENTR EXIT
ENTR accepts
the new setting
EXIT ignores the
new setting
KEY FUNCTION
<CH
Moves the cursor one character to the left.
CH>
Moves the cursor one character to the right.
INS
Inserts a character before the cursor location.
DEL
Deletes a character at the cursor location.
[?]
Press this key to cycle through the range of
numerals and characters available for
insertion. 0-9, A-Z, space ’ ~ ! # $ % ^ & * (
) - _ = +[ ] { } < >\ | ; : , . / ?
ENTR
Accepts the new setting and returns to the
previous menu.
EXIT
Ignores the new setting and returns to the
previous menu.
Some keys only appear as needed.
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE=500.0 PPB O3= XXXX
<TST TST> CAL SETUP
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
138 04315 Rev. C1