M400E Ozone Analyzer Operator’s Manual Specifications, Approvals and Warranty
2.2. EPA EQUIVALENCY DESIGNATION
Advanced Pollution Instrumentation, Inc., Model 400E photometric ozone analyzer is designated as Equivalent
Method Number EQOA-0992-087 as defined in 40 CFR Part 53, when operated under the following conditions:
Range: Any range from 100 ppb to 1 ppm.
Ambient temperature range of 5 to 40ºC.
Line voltage range of 105 – 125 VAC or 200 – 240 VAC, 50/60 Hz.
With 5-micron PTFE filter element installed in the internal filter assembly.
Sample flow of 800 ± 80 cc
3
/min at sea level.
Gas flow supplied by Internal or External pump.
Following Software Setting:
Table 2-4: Software Settings for EPA Equivalence
Dilution Factor 1.0
AutoCal ON or OFF
Dynamic Zero ON or OFF
Dynamic Span OFF
Dual range ON or OFF
Auto range ON or OFF
Temp/Pres compensation ON
Under the designation, the Analyzer may be operated with or without the following options:
Rack mount with slides.
Rack mount without slides, ears only.
Zero/Span Valves option.
Internal Zero/Span (IZS) generator.
4-20mA, isolated output.
2.3. CE MARK COMPLIANCE
EMISSIONS COMPLIANCE
The Teledyne Instruments Model 400E photometric ozone analyzer was tested and found to be fully compliant
with:
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B section 15.107 Class A, ICES-003 Class A (ANSI
C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.
Tested on 7/1/2002 to 7/2/2002 at CKC Laboratories, Inc., Report Number CE02-128.
SAFETY COMPLIANCE
The Teledyne Instruments Model 400E photometric ozone analyzer was tested and found to be fully compliant
with:
IEC 61010-1:90 + A1:92 + A2:95,
Tested in 08/2002 to 09/2002 at NEMKO: Report Number 2002-10280.
04315 Rev. C1 9