357BRemote Operation Of The M400E M400E Ozone Analyzer Operator’s Manual
8.5.6.2. Editing or Adding HESSEN Gas List Entries
To add or edit an entry to the Hessen Gas List, press:
SETUP X.X GAS ID:[ID Number]
000 ENTREXIT
SETUP X.X HESSEN GAS LIST
<SET SET> EDIT EXIT
SETUP X.X COMMUNICATIONS MENU
ID HESN COM1 COM2 EXIT
SETUP X.X HESSEN VARIATION:TYPE1
<SET SET> EDIT EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE=500.0 PPB O3= XXXX
<TST TST> CAL SETUP
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
Toggle these keys to
set the GAS ID to 400.
For new list entries
this number will be
displayed as
000.
Continue pressing NEXT until ...
EXIT discards the
new setting
ENTR accepts the
new setting
SETUP X.X O3, 0, 400, REPORTED
PREV MEXT INS DEL EDIT PRNT EXIT
SETUP X.X GAS TYPE:03
ENTR EXIT
There is only one GAS
TYPE available on
the M400E: O3
EXIT sets the
Gas typ too
NONE
SETUP X.X CONC RANGE:0
0ENTREXIT
Toggle this key to
set the concentration
range for the list
entry
SETUP X.X REPORTED:ON
ON ENTR EXIT
Toggle this key turn
ON/OFF the
REPORT attribute
SETUP X.X O3, 0, 400, REPORTED
PREV MEXT INS DEL EDIT PRNT EXIT
144 04315 Rev. C1