Teledyne 400E Dollhouse User Manual


 
M400E Ozone Analyzer Operator’s Manual 346BEPA Protocol Calibration
Table 10-1: Daily Activity Matrix
CHARACTERISTIC ACCEPTANCE LIMITS
FREQUENCY AND
METHOD OF
MEASUREMENT
ACTION IF
REQUIREMENTS ARE
NOT MET
Shelter
Temperature
Mean temperature between 22C
and 28C (72F and 82F), daily
fluctuations not greater than
2C.
Check thermograph chart
daily for variations not
greater than 2C (4F).
Mark strip chart for the
affected time period.
Repair/adjust temp control.
Sample
Introduction
System
No moisture, foreign material,
leaks, obstructions; sample line
connected to manifold.
Weekly visual inspection.
Clean, repair or replace as
needed.
Recorder
Adequate ink supply and chart
paper.
Legible ink traces.
Correct settings of chart speed
and range switches.
Correct time.
Weekly visual inspection.
Replenish and chart paper
supply
Adjust recorder time to
agree with clock note on
chart.
Analyzer
Operational
Settings
Flow and regulator indicators at
proper settings.
Temperate indicators cycling or
at proper levels.
Analyzer in sample mode.
Zero/span controls locked.
Weekly visual inspection. Adjust or repair as needed.
Analyzer
Operational Check
Zero and span within tolerance
limits as described in Subsec.
9.1.3 of Sec. 2.0.9
(Ref. 11).
Level 1 zero and span
every 2 weeks; Level 2
between Level 1 checks
at frequency desired by
user.
Isolate source error, and
repair.
After corrective action,
recalibrate analyzer.
Precision Check
Assess precision as described in
Sec. 2.0.8 (Ref. 11).
Every 2 weeks,
Sec. 2.0.8 (Ref. 11).
Calculate, report precision,
Sec. 2.0.8 (Ref. 11).
Table 10-2: Activity Matrix for Audit Procedure
AUDIT ACCEPTANCE LIMITS
FREQUENCY AND
METHOD OF
MEASUREMENT
ACTION IF REQUIREMENTS
ARE NOT MET
Multipoint calibration
audit
The difference between the
measured and the audit
values as a measure of
accuracy (Sec. 2.0.8 of Ref.
11).
At least once a quarter
(Sec. 2.0.8 of Ref. 11)
Re-calibrate the analyzer.
Data processing audit
Adhere to stepwise
procedure for data reduction
(Sec. 8.4); no difference
exceeding
0.02 ppm.
Perform independent
check on a sample of
recorded data, e.g.,
1 day out of every 2
weeks of data, 2 hours for
each day.
Check all remaining data if
one or more audit checks
exceeds 0.02 ppm.
Systems audit
Method described in this
section of the Handbook.
At the startup of a new
monitoring system, and
periodically as
appropriate; observation
and checklist.
Initiate improved methods
and/or training programs.
04315 Rev. C1 183