3.5.4. FUNCTIONAL CHECK
After the analyzer’s components have warmed up for at least 30 minutes, verify that the software properly
supports any hardware options that are installed.
For information on navigating through the analyzer’s software menus, see the menu trees described in Appendix
A.1.
Check to make sure that the analyzer is functioning within allowable operating parameters.
Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as their
expected values.
These functions are also useful tools for diagnosing problems with your analyzer (Section 13.1.2).
The enclosed Final Test and Validation Data sheet (part number 04314) lists these values before the
instrument left the factory.
To view the current values of these parameters press the following key sequence on the analyzer’s front panel.
Remember until the unit has completed its warm up these parameters may not have stabilized.
5. If your analyzer has an Ethernet card (Option 63) installed and your network is running a dynamic host
configuration protocol (DHCP) software package, the Ethernet option will automatically configure its
interface with your LAN. However, it is a good idea to check these settings to make sure that the DHCP
has successfully downloaded the appropriate network settings from your network server (See Section
8.4).
If your network is not runni
ng DHCP, you will have to configure the analyzer’s interface manually (See
Section 8.4.2.1).
04315 Rev. C1 29