M400E Ozone Analyzer Operator’s Manual Advanced FEATURES of the M400E analyzer
To access the ANALOG I/O CONFIGURATION sub menu, press:
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE=500.0 PPB O3= XXXX
<TST TST> CAL SETUP
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTR EXIT
AIO Configuration Submenu
DIAG SIGNAL I/O
NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO A OUTS CALIBRATED: NO
<SET SET> CAL EXIT
DIAG AIO CONC_OUT_1: 5V, OVR, NOCAL
<SET SET> EDIT EXIT
DIAG AIO AIN CALIBRATED: NO
<SET SET> CAL EXIT
DIAG AIO CONC_OUT_2: 5V, OVR, NOCAL
<SET SET> EDIT EXIT
DIAG AIO TEST_OUTPUT: 5V,OVR, NOCAL
<SET SET> EDIT EXIT
Toggle these
keys to enter the
correct
PASSWORD
Adjusts the signal output
for Analog Output A1
Adjusts the signal output
for Analog Output A2
Selects the parameter top be
output on the TEST channel
and adjusts its signal output
Figure 7-4: Accessing the Analog I/O Configuration Submenus
04315 Rev. C1 99