M400E Ozone Analyzer Operator’s Manual 357BRemote Operation Of The M400E
8.2. REMOTE ACCESS BY MODEM
The M400E can be connected to a modem for remote access. This requires a cable between the analyzer’s
COM port and the modem, typically a DB-9F to DB-25M cable (available from Teledyne Instruments with part
number WR0000024).
Once the cable has been connected, check to make sure:
The DTE-DCE is in the DCE position.
The M400E COM port is set for a baud rate that is compatible with the modem,
The Modem is designed to operate with an 8-bit word length with one stop bit.
The
MODEM ENABLE communication mode is turned ON (Mode 64, see Section 8.1.4).
Once this is completed, the appropriate setup command line for your modem can be entered into the analyzer.
The default setting for this feature is
AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0
This string can be altered to match your modem’s initialization and can be up to 100 characters long.
To change this setting press:
SETUP X.X COMMUNICATIONS MENU
ID COM1 COM2 EXIT
SETUP X.X COM1 MODE:0
<SET SET> EDIT EXIT
SETUP X.X COM1 PORT INIT:AT Y0 &DO &H &I0
<SET SET> EDIT EXIT
Continue pressing <SET or SET> until ...
SETUP X.X COM1 PORT INIT:AT Y0 &DO &H &I0
<CH CH> INS DEL [A] ENTR EXIT
EXIT discards the
new setting
ENTR accepts the
new setting
The <CH and CH>
keys move the cursor
left and right along the
text string
The INS and CH> key
inserts a new
character before the
cursor position
The DEL
deletes
character at
the cursor
position
Toggle this key to cycle through the
available character set:
Alpha: A-Z (Upper and Lower
Case);
Special Characters: space ’ ~ ! # $
% ^ & * ( ) - _ = +[ ] { } < > | ; : , . / ?
Numerals: 0-9
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE=500.0 PPB O3= XXXX
<TST TST> CAL SETUP
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
04315 Rev. C1 127