M400E Ozone Analyzer Operator’s Manual Advanced FEATURES of the M400E analyzer
7.4.2.2. Automatic Calibration of the Analog Outputs
To calibrate the outputs as a group with the AOUTS CALIBRATION command, select the ANALOG I/O
CONFIGURATION
submenu (see Figure 7-4) then press:
NOTE
Before performing this procedure, make sure that the AUTO CAL for each analog output (See Section
7.4.2.1)
NOTE:
Manual calibration should be used for any analog output set for a 0.1V output range or in cases where
the outputs must be closely matched to the characteristics of the recording device.
04315 Rev. C1 101