When viewing the Heater window, the typical heater configuration is as follows:
Heater 1 is the analytical block heater.
Heater 2 is the “high hat” heater.
The Temperature column on the Heaters window displays the current temperature; the
Current PWM column displays the percentage of power being used to run the heater.
The settings and values shown in the Heaters window and described in the table below are
preset at the factory and are based on the specific customer application. These values
should not be changed unless recommended by Application Engineering, Customer
Service personnel, or as part of a factory application requirement.
Function Typical Setting
Detector(s) or analytical block temperature 80 °C (176 °F)
Oven temperature 80 °C (176 °F)
Spare
Or, Methanator
Or, LSIV
N/A
300 °C (572 °F)
150 °C (302 °F)
FID configuration
When connected to the GC via MON2020, select Detectors from the Hardware menu to
access the Detectors dialog. Refer to the MON2020 user manual for additional
configuration details.