Emerson 700 XA Card Game User Manual


 
Alarm Name Possible Causes/Solution
Stream 1 Validation Failure
Stream 2 Validation Failure
Stream 3 Validation Failure
Stream 4 Validation Failure
Stream 5 Validation Failure
Stream 6 Validation Failure
Stream 7 Validation Failure
Stream 8 Validation Failure
Stream 9 Validation Failure
Stream 10 Validation Failure
Stream 11 Validation Failure
Stream 12 Validation Failure
Stream 13 Validation Failure
Stream 14 Validation Failure
Stream 15 Validation Failure
Stream 16 Validation Failure
Stream 17 Validation Failure
Stream 18 Validation Failure
Stream 19 Validation Failure
Stream 20 Validation Failure
The most recent validation sequence for the indicated stream
failed.
Recommended actions:
1. Check that the validation gas cylinder isolation valves are
open.
2. Check that the validation gas regulators are set properly.
3. If the validation gas regulator is below the set point, replace
the gas bottle with a full one.
4. If the gas used for validation is the same as the gas that is
used for calibration, ensure that the cylinder gas composition
value listed on the cylinder's tag or on the certificate of analy-
sis received from the supplier matches the value displayed in
MON2020's Component Data table.
5. Re-run the validation sequence.
6. If still unsuccessful contact your Emerson Process Manage-
ment representative.
Stream 1 RF Deviation
Stream 2 RF Deviation
Stream 3 RF Deviation
Stream 4 RF Deviation
Stream 5 RF Deviation
Stream 6 RF Deviation
Stream 7 RF Deviation
Stream 8 RF Deviation
Stream 9 RF Deviation
Stream 10 RF Deviation
Stream 11 RF Deviation
Stream 12 RF Deviation
Stream 13 RF Deviation
Stream 14 RF Deviation
Stream 15 RF Deviation
Stream 16 RF Deviation
Stream 17 RF Deviation
Stream 18 RF Deviation
Stream 19 RF Deviation
Stream 20 RF Deviation
The most recent calibration sequence failed.
Recommended actions:
1. Check that the calibration gas cylinder isolation valves are
open.
2. Check that the calibration gas regulators are set properly and
that the cylinder is not below the set point. If the cylinder is
below the set point, replace it with a full cylinder.
3. Verify that the calibration cylinder gas composition value lis-
ted on the cylinder tag or on the certificate of analysis re-
ceived from supplier matches the calibration cylinder gas
composition value displayed in MON2020's Component Data
table. If there is a mismatch, edit the Component Data table
to reflect the correct value. Re-run the calibration sequence.
4. If still unsuccessful contact your Emerson Process Manage-
ment representative.
Operation and maintenance
4
Operation and maintenance
85