The switch panel
The switch panel contains a network of on/off switches that allow you to manually control
the GC’s stream and analytical valves.
8-stream switch panel (left) and 18-stream switch panel (right)Figure 2-2:
There are two types of switch panels: 8-stream and 18-stream. The 8- stream switch panel
is the standard panel, and is used when the GC has only one heater/solenoid board
installed; if two heater/solenoid boards are installed, then the 18-stream switch panel is
used.
Valve switch from switch panel set to “OFF”Figure 2-3:
A valve has the following three operational modes:
AUTO - The valve turns on and off according to the Timed Events table that is
accessible through MON2020. To set a valve to AUTO mode, set its switch on the
switch panel to the “up” position.
Equipment description and specifications