Note
The serial number for MON2020 is located on the back of its CD-ROM case.
2. If your GC is configured with an FID, remove the vent plug from the FID outlet.
The vent plug has a tag attached to it that reads “REMOVE VENT PLUGS PRIOR TO
OPERATION”. Failure to remove the cap could result in a performance failure or in
damage to the detector.
Installation and startup should proceed only if all required materials are on hand and free
from obvious defects.
If any parts or assemblies appear to have been damaged in shipment, first file a claim
with the carrier. Next, complete a full report describing the nature and extent of the
damage and forward this report immediately to your Emerson Process Management
representative. Include the GC's model number in the report. Disposition instructions will
be provided as soon as possible. If you have any questions regarding the claim process,
contact your Emerson Process Management representative for assistance.
3.4.3 Required tools and components
You will need the following tools and components to install the 700XA:
Zero grade carrier gas:
- 99.995% pure
- Less than 5 ppm water
- Less than 0.5 ppm hydrocarbons
High pressure dual-stage regulator for the carrier gas cylinder:
- High side up to 3000 psig
- Gauge (psig)
- Low side capable of controlling pressure up to 150 psig
Calibration standard gas with correct number of components and concentrations.
Dual-stage regulator for the calibration gas cylinder with a low pressure side capable
of controlling pressure up to 30 psig.
Sample probe regulator (fixture for procuring the stream, or sample gas for
chromatographic analysis).
Coalescing filter.
Membrane filter.
Eighth-inch stainless steel tubing:
- For connecting calibration gas to the GC
- For connecting carrier gas to the GC.
- For connecting stream gas to the GC.
Heat tracing, as required, for sample transport and calibration lines.
Miscellaneous tube fittings, tubing benders, and tubing cutter.
Installation and setup
38