Emerson 700 XA Card Game User Manual


 
WARNING!
The unit is required to be used in a well ventilated area.
WARNING!
All gas connections must be properly leak tested at installation.
WARNING!
No user replaceable part inside except a few parts that are only allowed to be accessed by
trained service personnel.
WARNING!
Observe and comply with all precautionary signs posted on the GC. Failure to do so may result
in injury or death to personnel or cause damage to the equipment
WARNING!
If you plan to place the GC in a sealed shelter, always vent the GC to atmosphere with ¼-inch
tubing or larger. This will prevent the build up of H2 and sample gas.
WARNING!
Exit ports may discharge dangerous levels of toxic vapors; use proper protection and a suitable
exhaust device.
CAUTION!
Waste electrical and electronic products must not be disposed of with household waste. Please
recycle where facilities exist. Check with your local authority or retailer for recycling advice.
Note
The 700XA is CSA-certified and ATEX-certified. See the certification tag on the GC for specific details
about its agency approvals.
The following special conditions for safe use must be met:
Note
When installed, the equipment shall pass an electrical strength test of (1000 + 2 Un) V, rms applied
for at least 10 seconds up to a maximum of 60 seconds.