Emerson 700 XA Card Game User Manual


 
Type Specification
Analog inputs 2 connectors on the backplane, isolated
Standard analog
outputs
6 connectors on the backplane, isolated
Optional analog
outputs
8 connectors on optional expansion boards, isolated
Discrete digital in-
puts
5 connectors on the backplane
Discrete digital out-
puts
5 relay “Form C” contacts on Phoenix connectors on the backplane. Relay
contact rating 24 VDC nominal @ 1 Amp
Solenoid Driver
board
Maximum current output:
Channels 1 - 8: Maximum rating is 17 W. Each channel can drive 2.12 W or
88.4 mA at 24 VDC.
Channel 9 - 12: Maximum rating is 17 W. Each channel can drive 4.24 W or
176.8 mA at 24 VDC.
Transient protection Over-voltage category II
2.2.3 Airless analytical oven
The following table lists the specifications for the oven assembly.
Type Specification
Valves 6-port and 10-port XA valves; piston-operated diaphragms with pneumatic
actuation
Columns Max of 90 ft (27.4 m) of micro-packed columns; 1/16-inch
(1.6-mm) outside diameter
or
300 ft (91.4 m) of capillary columns
Solenoid actuation 24 VDC
Max 120 psi
Temperature con-
trol
24 VDC
2 heaters
2 optional heaters
Max oven operating temperature of 150° C (302° F)
2.2.4 Software
The following table lists the specifications for the GC’s software.
Type Specification
Software Windows-based MON2020.
Equipment description and specifications
2
Equipment description and specifications
23