Type Specification
Firmware Embedded firmware. Can be upgrade with MON2020.
Methods 4 Timed Event tables and 4 Component Data tables assignable to each stream.
Peak Integration • Fixed time or auto slope and peak identification.
• Update retention time upon calibration or during analysis.
2.2.5 Corrosion protection
Type Specification
Enclosure Material Copper-free and aluminum-coated with indus-
trial grade powder coat suitable for high humidi-
ty and salt-laden environments.
Process Wetted Materials Stainless steel; if the function of an item ex-
cludes the use of stainless steel, such as the
glass rotameter tubes, materials that are resist-
ant to corrosion are used.
Electronics All electronic circuit boards are tropicalized with
a clear conformal coating.
2.2.6 Archived Data Storage Capabilities
Type
Maximum Number of Re-
cords Remarks
Analysis Results 31744 88 days with 4 minute cycle
time
Final Calibration Results 370 1 year of Final Calibration re-
sults
Calibration Results 100
Final Validation Results 370 1 year of Final Validation results
Validation Results 100
Analysis Chromatograms 1703 Approximately 4.5 days assum-
ing 4 minute cycle time
Final Calibration Chromato-
grams
370 1 year of Final Calibration Chro-
matograms*
Final Validation Chromato-
grams
370 1 year of Final Validation Chro-
matograms*
Protected Chromatograms 100 User -selectable
Hourly Averages (Up to 250**
variables)
2400 100 days
Equipment description and specifications
24