4.4.11 Measure vent flow
You will need an accurate flow meter for this measurement.
To measure the measure vent flow, do the following:
Measure flow ventsFigure 4-16:
1. Consult the documentation the Parameter List that was provided with the GC to
learn the appropriate flow rate.
2. Attach a flow meter to the vent output on the right side of the GC that is labelled
“MV1”. The flow should match the value displayed in the Parameter List.
3. Attach a flow meter to the vent that is labelled “MV2”. The flow should match the
value displayed in the Parameter List.
4.4.12 Electrical components
The GC is designed to operate for long periods of time without the need for preventive or
regularly scheduled maintenance. It was designed with an explosion-proof enclosure that
is dust-proof, water-proof, and flame-proof.
WARNING!
Disconnect all electrical power to the unit and ensure the area is free of explosive gases. Failure
to follow this warning may result in injury or death to personnel or cause damage to the
equipment.
Prior to opening the GC, use MON2020 to ensure that there are no configuration or
parameter errors.