Tests and Error Messages
70 Chapter 7 Tests and Error Messages
Permanent Test
Note This is similar to the Cyclic Test described on page 35.
To carry out the Permanent Test:
1. Disconnect all transducers from the monitor and disconnect or switch off Telemetry.
2. While pressing press
Test:
C01 is shown in the US1/US display.
0 or 1 is shown in the Toco display.
3. Press again:
A01 is shown in the US1/US display.
4. Press the key twice:
A03 is shown in the US1/US display.
5. For each signal to be tested, connect a transducer to the corresponding socket within
15 seconds: US1/US, Toco, US2, and/or ECG.
6. Connect the Remote Event Marker.
7. While pressing the button on the marker, press the Toco Baseline Key. The test runs
for as long as the button on the marker is pressed.
FSpO
2
There are no operator error message for FSpO
2
. If no FSpO
2
annotation is printed on the
trace, you should refer to the troubleshooting flowchart -93.