Configuring the Monitor Using a PC
40 Chapter 4 Configuring the Monitor
Clearing the Error Log
To clear the error log, complete the following steps:
1. Disconnect all transducers from the monitor and disconnect or switch off Telemetry.
2. Press and
Test at the same time:
C01 is shown in the US1/US display.
0 or 1 is shown in the Toco display.
3. Press again to select the Function Menu:
A01 (Print the Error Log) is shown in the US1/US display.
4. Press to select
A02 (Clear the Error Log).
5. Press to clear the error log.
The confirmation message
Are You Sure? asks if you do want to clear the error log:
Press the Y key to clear the error log
or
Press the N key.
You then return to the Service Menu.
Series 50 Fetal Monitor Configuration and Service Software Rev A.nn.nn
Service Menu
Cyclic Test
Read options
Read ErrorLog
Clear ErrorLog Are you sure?
Write Ser.nr./Set Options
Fetal Heart Rate Alerting
Return to Main Menu
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
Serial Number : 2948G00010
Software Revision : A.02.03