Configuring the Monitor Using Pushbuttons
Chapter 4 Configuring the Monitor 27
Table 4-2Configuration Options
Menu Setting Options Default
C01 Time Format 0=AM/PM
1=24-hour
0
C02 Date Format 0=Month/Day/Year
1=Day.Month.Year
0
C03 IUP Format 0=mmHg
1=kPa
0
C04 Paper Format 0=US (30-240)
1=Europe (50-210)
0
C05
Recorder Print Offset
1.
0 to 11
0=right 1=left
0
C06 Recorder Heat Adjust 0 to 11
0=minimum 11=maximum
11
C07 Language Option 1=International
2=French
10=Japanese
13=Chinese simplified
1
C08 System Alert acknowledge at Marker
button
0=off
1=on
0
C09 Note Output to System 0=off
1=on (Roman 8).
1
C10
Combined Interface Module Settings
2
0=Barcode Reader
1=Dinamap 1846/8100 NIBP Monitor
2=COLIN Model BP-8800 NIBP Monitor
3=Nellcor OxiFirst FSpO
2
Monitor (N-400)
0
C11 External Toco gain 0=100% External Toco gain
1=50%
0
C12 NST Timer,
Paper-Out-Alert
3.
See Table 4-3 4
C15 Remote Patient Data after phone
transmission
0=Clear remote patient data
1=Keep remote patient data
0
A01
Print Error Log
4.
A02
Clear Error Log
4.
A03
Start Permanent Test
4.
1. To find the correct setting, connect a resting Toco transducer (one that is not under any load) to the monitor and then change the
setting until the trace is recording 20 units on the paper. Because of the 15-second time-out feature, and the delay between chang-
ing the setting and seeing the change on the paper, you may have to repeat this procedure to set the offset.
2. The interface to an NIBP monitor is only available on the Combined Interface Module with the label M1353-66531,E.
3. The NST auto recorder off option switches the recorder off automatically (advancing the paper to the next perforation) once the
NST time set has passed.
4. See -38 for more information on these menu items.