Philips M1353A Baby Accessories User Manual


 
Thermal Printhead
Chapter 11 Replacing Parts 151
Caution
The use of recorder paper that is not approved by Philips can damage the
Monitor. This type of damage will not be covered by warranty.
Note For more detailed illustrations of the enhancements to the recorder, see
Figure Figure 11-5, “Changes to the Recorder on page 125.
Thermal Printhead
Before you can remove the thermal printhead, you must first remove the top cover,
Recorder Board, and the recorder chassis:
1. Remove the top cover and the Recorder Board.
2. Place the cover with the paper table facing upwards and slide back the drawer
assembly.
3. Remove the two countersunk screws (S2 Figure 11-13).
4. Turn over the cover and place it so that the stepper motor is on the left and facing
upwards.
5. Remove the RFI bracket:
a. Remove the two screws (S1 in Figure 11-13).
b. Prise up and lift out the RFI bracket (it is fitted with a lug).
6. Remove the recorder chassis:
a. Remove the two screws (S3 in Figure 11-13).
b. Lift out the recorder chassis.