Safety Testing
102 Chapter 9 Preventive Maintenance, Care and Cleaning
When to
Perform Safety
Tests
Table 9-2M1351A and M1353A: When to perform safety test blocks
Service Event Test Block(s) Required
Installation
The product is customer installed.
You will find installation instructions in this
Installations and Service Guide.
Perform Visual, Power On and Performance
test blocks (See Table 9-3).
Preventive Maintenance
Preventive Maintenance is the responsibility of
the customer
For preventive maintenance refer to page 96.
Perform Visual test block (see Table 9-3).
Repair
This Installation and Service Guide contains
repair instructions for the 50 A and 50 IP moni-
tors.
Perform Visual, Power On and Performance
test blocks (see Table 9-3),
when power switch is replaced perform S2-Safety
test block (see page 104),
when transformer is replaced perform S3-Safety
test block (see Table 9-3).
When Frontend-Board is replaced perform S3-
Safety test block (see Table 9-3).
Upgrade
This Installation and Service Guide contains
upgrade instructions for the 50 A and 50 IP
monitors.
Perform Visual, Power On, Performance and
Safety test blocks (see Table 9-3).
All other Service Events Perform Visual, Power On and Performance