172 Appendix B Safety and Environment
The Monitor is designed to fulfill safety requirements according to IEC 60601-1, UL 544
and CSA C22.2-601.1-M90. It is not intended to be used for direct cardiac application and
it is not defibrillator-protected.
The external Toco transducer (M1355A) is equipment type B .
M1351A: Connecting the transducer to the Toco channel results in B
condition.
M1353A: Connecting the transducer to the Toco channel results in CF
condition.
The IUP pressure transducer (CPJ840J5) IUP quartz transducer (1290A and the 1290C)
are equipment type CF . Connecting the transducers to the Toco channel results in CF
condition.
The ultrasound transducer (M1356A) is equipment type B . Connecting the
transducer to the US channel results in B condition.
M1353A: The DECG transducer (M1357A) and MECG transducer
(M1359A) are equipment type CF . Connecting the
transducer to the ECG channel results in CF condition.
Environment
Use the monitor in an environment that is reasonably free from vibration, dust, corrosive
or explosive gases, flammable agents, extremes of temperature, humidity and so forth. It
operates within specifications at ambient temperatures between 0 and 55
o
C. Ambient
temperatures that exceed these limits can affect the accuracy of the monitor and cause
damage to the components and circuits. Only products that fulfil the necessary safety and
electrical standards should be used in conjunction with the monitor (contact your local
response center for details).
Allow at least 5cm (2in) clearance around the monitor for proper air circulation. If the
monitor is mounted in a cabinet, allow sufficient space at the front for operation and at the
rear for servicing with the cabinet door open.