RS232 Serial Interface
Chapter 10 Peripherals 113
FSpO
2
Monitor If the fetal monitor has the Combined Interface Module J-14 fitted, you can interface to an
FSpO
2
monitor (Nellcor N-400 or other compatible). You must have software revision
number A.02.09 or higher.
To connect an FSpO
2
monitor to the M1351A or M1353A:
1. Use interface cable M1353-61614. This has identical 9-pin male connectors at each
end.
2. Connect one end to the M1351A or M1353A and the other to the FSpO
2
monitor.
3. Secure the interface cable by the screws at the cable connectors.
Figure 10-7FSpO
2
Interface Cable (M1353-61614)
Baudrates The baudrates for the FSpO
2
monitor must be set as follows:
Baudrate: 2400
Startbit: 1 Parity: None
Databits: 8
Stopbits: 1
Cable: M1353-61614
Table 10-3 NIBP Monitor and Minimum Time Interval
Paper Speed Minimum Time Interval
1 cm/min 3 minutes.
2 cm/min 2 minutes.
3cm/min 1 minute.