Philips M1353A Baby Accessories User Manual


 
Configuring the Monitor Using Pushbuttons
28 Chapter 4 Configuring the Monitor
Note After you put paper into the recorder, let it run for a few pages (use the
advance key) to allow the paper to align itself correctly to the right side.
Examples: How
to Change the
Time Format
and IUP Format
using
Pushbuttons
Example 1:
To change the time format from AM/PM format to 24-hour format, do the following:
1. Disconnect all transducers from the monitor and disconnect or switch off Telemetry.
2. While pressing , press
Test.
3. Press the key to change the setting to
1.
4. Press
Test to store the new settings.
Example 2:
To change the IUP format from mmHg to kPa, do the following:
Disconnect all transducers from the monitor and disconnect or switch off Telemetry.
While pressing , press
Test.
Press repeatedly until
C03 is displayed.
Press the key to change the setting to
1.
Press
Test to store the new settings.
Table 4-3 NST Timer Paper-Out-Alert
Options NST Timer
NST automatic
recorder off
Paper-Out-Alert
0
✗✗✗
1
✓✗ ✗
2
✓✗ ✗
3
✗✗✓
4
✓✗✓
5
✓✓