Philips M1353A Baby Accessories User Manual


 
Configuring the Monitor Using a PC
Chapter 4 Configuring the Monitor 35
Service Menu
To select an item from the menu, move the cursor to the item you require and press
Enter.
However, before you select
Cyclic Test, connect ultrasound, Toco and DECG
transducers to the monitor.
Note The serial number and software revision of your monitor may be different
from those shown.
Running the Cyclic Test
This runs a continuous Self Test. Any errors are written to the Error Log and can be read
using
Read ErrorLog. Press any key to stop the test and return to the Service Menu.
Note During the test, the keys on the front panel are disabled.
Series 50 Fetal Monitor Configuration and Service Software Rev A.nn.nn
Service Menu
Cyclic Test
Read options
Read ErrorLog
Clear ErrorLog
Write Ser.nr./Set Options
Fetal Heart Rate Alerting
Return to Main Menu
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
Serial Number : 2948G00010
Software Revision : A.02.03
Series 50 Fetal Monitor Configuration and Service Software Rev A.nn.nn
MESSAGE:
CYCLIC TEST is started
It starts a continuous self test of the monitor
Press any key to stop this test
Read ErrorLog to see the errors that occurred while self testing