Philips M1353A Baby Accessories User Manual


 
RS232 Serial Interface
Chapter 10 Peripherals 111
NIBP Monitor One of the following devices can be connected to an M1351A or M1353A to record
maternal NIBP:
Dinamap1846/8100 NIBP Monitor. This requires a Critikon model 8801 adapter
COLIN Press-Mate/Nippon Colin Listmini Model BP-8800.
Before connecting a NIBP monitor, ensure that the correct configuration options on the
M1351A and M1353A have been selected. See Table 4-2 on page 27 for the options
available. Interface to an NIBP Monitor is available only on the Combined Interface
Module, Option J13, with the label M1353-66531, E (software revision number A.02.xx or
higher - the software revision number is written on the error log).
Connecting
an NIBP Monitor to
the M1351A or
M1353A
To connect a Dinamap 1846/8100 NIBP Monitor to the M1351A or M1353A:
1. Use the interface cable M1350-61609.
2. Connect the 9-pin end of the cable to the fetal monitor.
3. Connect the 25-pin end of the cable to the external device.
4. Secure the interface cable by the screws at the cable connectors.
Figure 10-5Interface Cable M1350-61609
Caution
Before connecting an NIBP Monitor, use the Grounding Cable (part number
8120-2961 for USA or 8120-4808 for Europe) to connect the equipotential
grounding point on the fetal monitor to earth potential. The power cord of
the NIBP Monitor must be plugged into a wall-mounted power outlet - not
into an extension block.