Fuses
136 Chapter 11 Replacing Parts
Fuses
The fuse values are printed beside the mains socket:
To replace the fuses:
1. Switch off the monitor and disconnect it from the main power supply.
2. Using a flat-bladed screwdriver, prise open the fuse cover.
3. Lift the fuse holder slightly and pull it out.
4. Remove the fuse from the holder and replace it with another of the correct value.
5. Slide the holder back into place, aligning the arrow on the holder with the arrow on
the cover.
6. Repeat steps 3, 4 and 5 for the second fuse.
7. Close the fuse cover.
For 100-120V Line Voltage T500mA/250V
For 220-240V Line Voltage T250mA/250V