Power Supply Board
140 Chapter 11 Replacing Parts
Power Supply Board
To remove the power supply board:
1. Remove the top cover.
2. Disconnect the cables (A and C).
3. Remove the three screws (S). The S2 screws are longer than the S1 screws.
4. Lift out the board.
To refit the board, reverse the above procedure. After refitting the board, carry out the
Parameter Test (see page 68).
Table 11-16Fuse Part Numbers
All Fuses
630 mA 2110-0895