Error Log Messages
44 Chapter 4 Configuring the Monitor
Error Log Messages
Table 4-5Error Log Messages
Message Location of Error and Solution
Err 18 Incorrect type of paper loaded (same as Err PAP).
Replace with recommended paper.
Err 19 Battery low or empty of charge (same as Err bAt).
Change the batteries as soon as possible.
Err 70 Modem not responding.
Modem is either not installed, incorrectly installed, faulty or not an approved
model.
Err 77 Error during transmission of data.
Probably due to external sources. Retransmit the data.
Err 500 General failure - no specific location.
See the Troubleshooting Flowchart in Chapter 8.
Err 501,
513
Front End Board (Single Ultrasound model).
See the Troubleshooting Flowchart in Chapter 8.
Err 502 Power Supply Board.
See the Troubleshooting Flowchart in Chapter 8.
Err 503 CPU Board.
See the Troubleshooting Flowchart in Chapter 8.
Err 510 Recorder Board.
See the Troubleshooting Flowchart in Chapter 8.
Err 511 Front End Board (Dual Ultrasound model).
See the Troubleshooting Flowchart in Chapter 8.
Err 512 Intrapartum Board.
See the Troubleshooting Flowchart in Chapter 8.
Err 531 Combined Interface Board.
See the Troubleshooting Flowchart in Chapter 8.
Err 532 Modem Interface Board.
See the Troubleshooting Flowchart in Chapter 8.
Err 601 Paper feed.
See the Troubleshooting Flowchart in Chapter 8.
Err 602 Incorrect type of paper.
Load paper with the correct scale or change the setting.
Err 610 No loudspeaker, or fuse has blown.
See the Troubleshooting Flowchart in Chapter 8.