is restricted to Administrator users. You can start the upgrade by clicking Upgrade firmware in the instrument’s context-
sensitive menu or the [Upgrade Firmware] button in the instrument tab. When you start the upgrade process, the
instrument is logically disconnected from the system and the instrument icon will disappear from the status bar.
4. A black UpgradeIt command window opens.
Firmware Installation Folder
5. If you are upgrading the firmware for the first time using this computer, you must minimize the UpgradeIt window at this
time and install the supporting device drivers.
6. The automatic installation will fail. Open the Device Manager, right-click Device Firmware Upgrade, and select Upgrade
Driver Software.
7. Select Browse my computer for driver software.
8. Click Browse.
9. Browse to the Dfu\Drivers folder location in the Molecular Detection System folder described in step 2 above.
10. A Windows Security window will appear. Click Install this driver software anyway.
11. After the wizard installs the drivers, the computer completes its device recognition process and notifies you that the new
device is ready for use.
Device Manager