Well Result Symbol Result Interpretation
Sample Positive
The sample is presumptive positive
for the target pathogen.
Sample Negative The sample is negative for the target pathogen.
Sample Inhibited
The sample matrix was inhibitory to the assay. A
re-test may be required. Refer to the troubleshooting
section and the assay kit Product Instructions for more
information.
Sample Inspect
The presence or absence of the target pathogen was
indeterminate. A re-test may be required. Refer to
the troubleshooting section and the assay kit Product
Instructions for more information.
Sample Error
No bioluminescence was detected. A re-test may be
required. Refer to the troubleshooting section and the
assay kit Product Instructions for more information.
Reagent
Control
Valid The Reagent Control was valid.
Reagent
Control
Invalid
The Reagent Control was invalid. A re-test may be
required. Refer to the troubleshooting section and the
assay kit Product Instructions for more information.
Reagent
Control
Inspect
The Reagent Control was indeterminate. A re-test may
be required. Refer to the troubleshooting section and
the assay kit Product Instructions for more information.
Reagent
Control
Error
No bioluminescence was detected. A re-test may be
required. Refer to the troubleshooting section and the
assay kit Product Instructions for more information.
Negative
Control
Valid The Negative Control was valid.
Negative
Control
Invalid
The Negative Control was invalid. A re-test may be
required. Refer to the troubleshooting section and the
assay kit Product Instructions for more information.
Negative
Control
Inspect
The Negative Control was indeterminate. A re-test may
be required. Refer to the troubleshooting section and
the assay kit Product Instructions for more information.
Negative
Control
Error
No bioluminescence was detected. A re-test may be
required. Refer to the troubleshooting section and the
assay kit Product Instructions for more information.
Matrix
Control
Valid The Matrix Control was valid.
Matrix
Control
Inhibited
The sample matrix was inhibitory to the Matrix Control.
A re-test may be required. Refer to the troubleshooting
section and the assay kit Product Instructions for more
information.
Matrix
Control
Error
No bioluminescence was detected. A re-test may be
required. Refer to the troubleshooting section and the
assay kit Product Instructions for more information.