Teledyne API Model 200AU NO
X
Analyzer Instruction Manual, 02293, Rev. F
6-4
Table 6-2: Example of AutoCal Setup
Step Action Comment
1. Press
SETUP-ACAL
This button sequence will cause the AUTOCAL menu to be
displayed.
2. Press PREV-NEXT Press PREV-NEXT until SEQ 2 is displayed
3. Press MODE Select the MODE menu
3. Press PREV-NEXT Press PREV NEXT to scroll to SPAN
4. Press ENTR ENTR selects the SPAN MODE
5. Press SET Select the SET menu to change the sequence attributes
6. Press PREV-NEXT Scroll the SET menu to TIMER ENABLE
7. Press EDIT Allows changing the TIMER ENABLE attribute, select ON
8. Press ENTR ENTR changes TIMER ENABLE to ON
9. Press PREV-NEXT Repeat steps 6-9 for each attribute
10. Press EXIT Press the EXIT key to return to upper level menus
6.4 4-20 mA, Current Loop Output
The current loop option replaces the voltage output of the instrument with an isolated 4-20 mA
current output. The current outputs come out on the same terminals that were used for voltage
outputs, see Figure 2-2. See Troubleshooting Section 9.3.3 for setup and calibration.
6.5 NO
y
Converter
The NO
y
Converter Option allows placement of the molybdenum converter at the sample inlet
point. This configuration is useful for background studies and where it is desirable to
immediately convert certain components of the atmosphere that decompose rapidly, or may be
lost on the walls of a normal sample induction system. This group of compounds is sometimes
referred to as NO
y
.
The Option consists of a modified M200AU, a M501Y chassis and remote converter. The
pneumatic diagram for this system of components is in Figure 8-7. Because of the nature of this
option, there is a separate manual “M501Y – Remotely Mounted Converter” (P/N 02808). The
manual covers setup, calibration, and operation of the system.