Teledyne 200AU Doll User Manual


 
Teledyne API Model 200AU NO
X
Analyzer Instruction Manual, 02293, Rev. F
7-28
Table 7-16: Activity Matrix for Data Quality
Characteristic Acceptance Limits
Frequency And Method
Of Measurement
Action If Requirements
Are Not Met
Shelter temperature Mean temperature
between 22
o
C and
28
o
C (72
o
and 82
o
F),
daily fluctuations not
greater than ±2
o
C
Check thermograph chart
weekly for variations
greater than ±2
o
C (4
o
F)
1. Mark strip chart for
the affected time
period
2. Repair or adjust
temperature control
Sample introduction
system
1. No moisture,
foreign material,
leaks, obstructions
2. Sample line
connected to
manifold
Weekly visual inspection Clean, repair, or replace
as needed
Recorder 1. Adequate ink &
paper
2. Legible ink traces
3. Correct chart
speed and range
4. Correct time
Weekly visual inspection 1. Replenish ink and
paper supply
2. Adjust time to agree
with clock; note on
chart
Analyzer operational
settings
1. TEST
measurements at
nominal values
2. M200AU in
SAMPLE mode
Weekly visual inspection Adjust or repair as
needed
Analyzer operational
check
Zero and span within
tolerance limits as
described in Subsec.
9.1.3 of Sec. 2.0.9
(Q.A. Handbook)
Level 1 zero/span every 2
weeks
Level 2 between Level 1
checks at frequency desired
by user
1. Find source of error
and repair
2. After corrective
action, re-calibrate
analyzer
Precision check Assess precision as
described in Sec. 2.0.8
and Subsec. 3.4.3
(Ibid.)
Every 2 weeks, Subsec.
3.4.3 (Ibid.)
Calc, report precision,
Sec. 2.0.8 (Ibid.)