Table 2-1: Final Test and Calibration Values (Continued)
Parameter
Observed
Value
Units Nominal Range Reference Section
NO Span Conc PPB 1 - 2000 Table 7-3
NO
x
Span Conc PPB 1 - 2000 Table 7-3
NO Slope -
1.0 ± 0.3
Table 7-18
NO
x
Slope -
1.0 ± 0.3
Table 7-18
NO Offset mV -10 to +150 Table 7-18
NO
x
Offset mV -10 to +150 Table 7-18
Moly Efficiency % 0.96 - 1.02 7.6.6, 5.2.2.6
Stability at Zero PPB 0.001 - 2000 Table 9-1
Stability at Span PPB < 1 @ 400 ppb Table 9-1
Measured Flows
Sample Flow cc/min 1000 ± 100 9.3.7, Figure 9-8
Ozone Flow cc/min 80 ± 15 9.3.7, Figure 9-8
Factory Installed Options Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Rack Mount, External Pump w/ Slides
Rack Mount, External Pump w/o Slides
Stainless Zero/Span Valves
Current Loop - NO
x
Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - NO Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - NO
2
Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - TST Chan 4-20 mA 0-20 mA Isolated Non-Isolated
PROM # Serial #
Date Technician