Analyzer Instruction Manual, 02293, Rev. F
5-35
Table 5-19: RS-232 Calibration Examples
Action RS-232 Commands Comments
Zero Calibration C ZERO
C COMPUTE ZERO
C EXIT
Z/S valves switched to admit
zero gas. Zero cal in Single
Range mode
Zero Calibration of low range -
AutoRange Enabled
C ZERO 1
C COMPUTE ZERO
C EXIT
Z/S valves switched to admit
zero gas. Zero calibration of
low range in Auto Range mode
Span Calibration of high range -
AutoRange Enabled
C SPAN 2
C COMPUTE SPAN
C EXIT
Z/S valves switched to admit
span gas. Span calibration of
high range in Auto Range
mode
Zero Calibration with Dynamic
Calibration enabled
C ZERO
C EXIT
Z/S valves switched to admit
zero gas. Instrument is zero
calibrated if DYN CAL is
enabled.
Zero Calibration C ZERO
C EXIT
Z/S valves switched to admit
zero gas. Instrument zero is
just checked, but not changed.
Execute AutoCal Sequence #2 C ASEQ 2 Execute a predefined AutoCal
Sequence. Executes sequence
immediately, ignoring time and
date parameters.
Span Calibration Check C SPAN
C EXIT
Z/S valves switched to admit
span gas. Instrument span is
just checked, but not changed.
Whenever the analyzer starts or finishes a ZS calibration, it issues a status report to the RS-232
interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise,
they will be discarded. Table 5-20 shows the format of the text of the calibration messages. An
example of an actual sequence of calibration status messages is:
C DDD:HH:MM IIII START MULTI-POINT CALIBRATION
C DDD:HH:MM IIII NOX=xxxxx PPB NO=xxxxx PPB NO2=xxxxx PPB
C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION