Teledyne 200AU Doll User Manual


 
Teledyne API Model 200AU NO
X
Analyzer Instruction Manual, 02293, Rev. F
5-19
Table 5-6: DAS Data Channel Editing
Step Action Comment
1.
Press SETUP-DAS-EDIT
Enter DAS menu to edit Data Channels
2.
Press PREV/NEXT
Select Data Channel to edit
3.
Press EDIT
Enter the Edit menu for the selected Data Channel
4.
Press SET> (5 times)
Scroll through Data Channel properties until RS-232
REPORT: OFF is displayed
5.
Press EDIT
Edit selected setup property
6.
Toggle OFF to ON
Change RS-232 REPORT property
7.
Press ENTR
Accepts change
8.
Press EXIT (4 times)
Exits back to sample menu
5.3.4 Range Menu
The instrument operates on any full scale range from 5 to 2000 ppb. The range is the
concentration value that equals the maximum voltage output on the rear panel of the instrument.
The front panel will read the concentration anywhere from 0 to 2000 ppb regardless of the range
selected.
NOTE
Only one of the following range choices can be active at any one time.
There are 3 range choices:
1. Single Range
2. Auto Range
3. Independent Ranges
5.3.4.1 Single Range
This range option selects a single range for all output channels (NO, NO
x
, NO
2
) of the M200AU.
To select Single Range press SETUP-RNGE-MODE-SING, then press ENTR. To set the value
for the range press SETUP-RNGE-SET, enter the full scale range desired from 5 ppb to 2000
ppb, then press ENTR.