Andersen Windows & Doors AN74IX Baby Accessories User Manual


 
9
OWNER’S MANUAL
SECTION 2
Diagrammatic Instructions for Normal Operation
Turn on the sterilizer by pressing the on-off switch on
the back right side of the top cabinet.
The Liquid Crystal Display will illuminate. The initial
startup and standby screen is shown to the right.
To load the sterilizer and commence a sterilization
cycle, press the button immediately to the right of the
word START.
If an abator is attached to the machine, the number of
cycles remaining for the abator cartridge will be dis-
played. A new abator cartridge may be used for a
maximum of 250 sterilization cycles.
The display will indicate that the sterilizer has
performed a self test on the cabinet ventilation pump
and ventilation switch. The version of the computer
software is shown (VER 4.04). In addition, the total
number of hours that the ventilation pump has been
operating is shown (35 PUMP HOURS). This infor-
mation is used to determine when preventive mainte-
nance should be performed. (See Section 7). The
cooling fan blade speed is displayed as a percentage of
nominal operating speed (FAN 97% RPM).
Once the self test of the cabinet ventilation pump has
been successfully completed, the display will show the
instructions for loading the sterilizer. After the sterili-
zation bag has been secured to the purge tube bobbin
via the Velcro strap, press the button to the right of the
word PURGE to evacuate air from the sterilization
bag.
The display will show the initial 1 minute, 30 second
initial purge is in progress. To stop the cycle and re-
turn to the standby screen, press the button to the left
of the word EXIT.
SELF TEST
OK
35 PUMP HOURS
FAN 97 % RPM
When PURGE pressed
INITIAL PURGE
00:01:30 REMAINING
EXIT
LOAD STERILIZER BAG
CLOSE BAG OVER TUBE
PURGE BAG
EXIT PURGE
Continued on next page
SELF TEST
VER 4.04
35 PUMP HOURS
FAN 97 % RPM
250 ABATOR CYCLES
REMAINING
When START is pressed
AN74 I
ANPROLENE STERILIZER
START
© Andersen Products Ltd