Flowchart for Technical and Diagnostic Screens
AN74 I
ANPROLENE STERILIZER
START
DIAGNOSTICS
NO ABATOR
DIAGNOSTICS
ABATOR ATTACHED
PURG V. 0.0 (>5.00)
VENT V. 0.00 (>0.25)
BACK P. 0.0 (<15.0)
EXIT FAN 102% (>70)
PURG V. 25.3 (>5.00)
VENT V. 0.00 (>0.25)
BACK P. 6.4 (<15.0)
EXIT ABATOR ON
PURG V. 0.0 (>5.00)
VENT V. 1.51 (>0.25)
BACK P. 8.2 (<15.0)
EXIT ABATOR OFF
PURG V. 25.3 (>5.00)
VENT V. 0.00 (>0.25)
BACK P. 6.4 (<15.0)
EXIT FAN 102% (>70)
If Abator
attached
If no Abator
attached
PURG V. 0.0 (>5.00)
VENT V. 0.94 (>0.25)
BACK P. 8.2 (<15.0)
EXIT FAN 102% (>70)
PRESS ABATOR TO
INSTALL CARTRIDGE,
OTHERWISE PRESS DIAG
ABATOR DIAG
PURG V. 0.0 (>5.00)
VENT V. 0.00 (>0.25)
BACK P. 0.0 (<15.0)
EXIT FAN 102% (>70)
These screens are used by authorized service technicians and installers only, to check the vacuum created
by the abator pump, ventilation pump and purge pump. The screens are used for diagnostic and installation
purposes and would not normally be used by the operator of the machine.
No pumps
running,
all sensors
should
equal 0.
Test Purge
pump
No pumps
running,
all sensors
should
equal 0.
Test
Purge
pump
Test Vent
pump