Andersen Windows & Doors AN74IX Baby Accessories User Manual


 
11
OWNER’S MANUAL
Diagrammatic Abator Instructions:
Installing Optional New Abator Cartridge, Resetting Counter
When the Abator cartridge has been used completely
(250 sterilization cycles), the screen will display
ABATOR CARTRIDGE EXPIRED and the alarm will
sound.
Additional cycles may be completed by bypassing the
screen. This is not
recommended. The cartridge
should be changed prior to starting a new cycle.
ABATOR CARTRIDGE
EXPIRED
EXIT BYPASS
Once a new cartridge is installed, the
counter must be reset.
To reset the counter:
1. Hold the left button continuously for
more than 3 seconds
2. Press ABATOR button
3. Answer YES to the question:
INSTALL NEW ABATOR
CARTRIDGE
When
left button
pressed and held
for more than 3
seconds.
AN74 I
ANPROLENE STERILIZER
PARAMETERS START
INSTALL NEW
ABATOR CARTRIDGE
NO YES
PRESS ABATOR TO
INSTALL CARTRIDGE,
OTHERWISE PRESS DIAG
ABATOR DIAG
CARTRIDGE COUNTER
RESET TO 250
DIAGNOSTICS
ABATOR ATTACHED