Press ALARM OFF
to stop the alarm
Flowchart for Ventilation Pump Failure During Sterilization Cycle
12 (or 24) HOUR CYCLE
STERILIZING
1:24:15 REMAINING
STERILIZATION HALTED
VENT PUMP FAILURE
1:24:15 REMAINING
ALARM OFF
STERILIZATION HALTED
VENT PUMP FAILURE
1:21:55 REMAINING
UNLOAD STERILIZER
STERILIZATION HALTED
VENT PUMP FAILURE
EXIT
If a VENT PUMP FAILURE occurs during
the sterilization cycle, the screen will display
the error message, and the machine will beep
continuously until the ALARM button is
pressed. This alerts the operator to the prob-
lem. The purge pump will start and remain on
for the remainder of the normal cycle time.
VENT PUMP FAILURE
REFER TO ERROR
SECTION OF MANUAL
In this case, sterilization has been stopped dur-
ing the cycle and the Ethylene oxide in the
liner bag was removed by the purge pump.
ITEMS IN THE LINER BAG SHOULD NOT
BE CONSIDERED STERILE
Refer to the Error and Troubleshooting sections
of the manual for information about repairing the
sterilizer
To reset the machine, shut off the machine power.
Then, while holding both
buttons, turn on the
power. Release buttons after the display shows
characters.
The machine may be unloaded once the
counter reaches 0:00:00, but the items in the
liner bag should be resterilized prior to use
AN74 I
ANPROLENE STERILIZER
START
RESET