3.13. At the end of the two hour ventilation cycle, the following screen will appear,
notifying the operator that the sterilizer may be unloaded.
UNLOAD STERILIZER
EXIT
3.14. If the operator does not unload the sterilization liner bag at the end of the two hour
ventilation cycle, the purge pump and the cabinet ventilation pump continue to alternate on a
two minute cycle to flush the sterilization liner bag and cabinet indefinitely until the door is
opened and the EXIT button is pressed This feature may be used for Additional Aeration.
Refer to page 45.
3.15. If the operator opens the door after the two hour purge cycle is complete and presses
EXIT , both pumps will stop and the program will automatically return the sterilizer to the
start screen.
3.16. To unload the sterilizer, open the door and remove the Velcro
®
strap from around the
neck of the purge tube. Remove the purge tube from the sterilization liner bag. Unload the
wrapped items from the sterilization liner bag and place them on a shelf in a well-ventilated
room that has a minimum of 10 air changes per hour of fresh makeup air.
3.18. The used gas release bag ampoule and sterilization liner bag may be disposed of in
ordinary rubbish. Never reuse sterilization liner bags.
3.17. Gas absorbent materials such as rubber or plastic must be aerated in their individual
packages for at least 24 additional hours before they are used.