Andersen Windows & Doors AN74IX Baby Accessories User Manual


 
66
OWNER’S MANUAL
12 (or 24) HOUR CYCLE
STERILIZING
1:24:15 REMAINING
VENTILATING BAG
PURGE PUMP FAILURE
24:00:00 REMAINING
UNLOAD STERILIZER
00:00:44
EXIT
VENTILATING BAG
0:23:55 REMAINING
If a PURGE PUMP FAILURE
occurs during the ventilation
cycle, an additional 24 hours
will automatically be added to
the time remaining
The PURGE PUMP FAILURE message will
remain on the screen. At the end of the ex-
tended cycle, the operator should: (1) open the
door; (2) remove the Velcro strap and open the
mouth of the bag; and (3) close and lock the
door.
Continue ventilation for an additional 2 hours.
Open the door and press EXIT. Then remove
the items from the cabinet as you normally
would.
ITEMS IN THE LINER BAG MAY BE
CONSIDERED STERILE
PURGE PUMP FAILURE
REFER TO ERROR
SECTION OF MANUAL
Flowchart for Purge Pump Failure During Ventilation Cycle
Refer to the Error and Troubleshooting sections
of the manual for information about repairing the
sterilizer
To reset the machine, shut off the machine power.
Then, while holding both buttons, turn on the
power. Release buttons after the display shows
characters.
AN74 I
ANPROLENE STERILIZER
START
PURGE PUMP FAILURE
OPEN DOOR
REMOVE VELCRO STRAP
CLOSE DOOR
RESET
PURGE PUMP FAILURE
01:59:59 REMAINING
Count up timer