Andersen Windows & Doors AN74IX Baby Accessories User Manual


 
61
OWNER’S MANUAL
At the end of the cycle,
the operator opens the
door and removes the
load. The operator then
closes the door and
presses the EXIT button.
This returns the steri-
lizer to the standby
state.
VENTILATING BAG
01:23:55 REMAINING
UNLOAD STERILIZER
00:18:54
EXIT
Additional aeration:
The purge pump and the venti-
lation pump continue running
alternate two minute cycles
until the door is opened and
the EXIT button is pressed.
Therefore, items may be left in
the machine for further aera-
tion after the cycle has ended.
During the ventilating bag cycle, the
purge pump and cabinet ventilation
pump run alternating two minute
cycles such that at least one pump is
running all the time, but two pumps
never run at the same time.
Flowchart for Normal Operating Screens, continued
AN74 I
ANPROLENE STERILIZER
START
Continued from previous page
After the initial
purge has been com-
pleted, the operator
breaks the ampoule
and closes the door.
If the operator does
not select the cycle
length within 5 sec-
onds, an alarm will
sound continuously
as a reminder.
BREAK AMPOULE
CLOSE & LOCK DOOR
SELECT CYCLE LENGTH
24 HOUR 12 HOUR
12 (or 24) HOUR CYCLE
STERILIZING
11:24:15 REMAINING
When cycle
length selected