[ Compliance verification and guidance ]
Guidance and manufacturer's declaration - electromagnetic emissions
The KOWA FM-600 is intended for use in the electromagnetic environment specified below. The customer or the user of
the KOWA FM-600 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Group 1
Class A
Class A
Complies
The KOWA FM-600 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The KOWA FM-600 is suitable for use in all establishments other than domes-
tic and those directly connected to the public low-voltage power supply net-
work that supplies buildings used for domestic purposes.
Guidance and manufacturer's declaration - electromagnetic immunity
The KOWA FM-600 is intended for use in the electromagnetic environment specified below. The customer or the user of the KOWA FM-
600 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast tran-
sient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
± 2, 4, 6kV contact
± 2, 4, 8kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 0.5,1 kV differential
mode
± 0.5,1,2 kV common
mode
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3 A/m
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commer-
cial or hospital environment.
Mains power quality should be that of a typical commer-
cial or hospital environment.
Mains power quality should be that of a typical commer-
cial or hospital environment.
If the user of the KOWA FM-600 requires continued op-
eration during power mains interruptions, it is recom-
mended that the KOWA FM-600 be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.
± 6kV contact
± 8kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV differential mode
± 2 kV common mode
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3 A/m
NOTE UT is the A.C. mains voltage prior to application of the test level.
13. Electromagnetic Compatibility