Kowa fm600 Inflatable Bouncy Toy User Manual


 
25
7-5 Recommended procedures when measured waveforms are offset to left or right
You may encounter waveform offset to the left as shown in Fig. 6 when
attempting a measurement at an excessively lower part of the anterior
chamber.
You can not obtain a correct flare value in this condition.
1 Select “ON” for “Laser : Position” as described in “8. Configuration and
settings” and change the laser position in BG check screen to solve this
problem.
2 Use “ ” and “ ” several times to change the laser position.
3 Press “Measurement switch” to perform a measurement after the position
change. This setting is effective until all measurement data for the one
examined eye are cleared
.
Select “End” in Summary data screen.
7. Tips for measurements
L
ListStop
Narrow
Wide
Fig. 6
Fig. 7