Stained lens may affect measurements.
It is recommended to calibrate regularly in order to obtain accurate measurements.
A message will appear during device start-up if the device had no calibration done for more than a month. Calibrate the device as
soon as possible when the message appears.
5. Calibration
Follow the calibration procedure described below.
It is recommended to calibrate the device in a dark room.Use a light shade (optional) when calibrating in a well-lighted room.
5-1 Calibration
1 While Frontal view alignment screen is displayed, turn Calibrator to
CALIBRATION position.
If the display shows “LD Temp check Please wait”, wait until the message disap-
pears.
If the display shows “Laser Attenuating …”, turn Calibrator until the message shown
in Fig. 3 appears in the display.
If the display shows “The acquired data is deleted. Is a calibration carried out ?”
indicating there is acquired data, select “Yes” to proceed to the calibration.
2 Slide the protective cap aside to expose the window (Fig. 2).
Caution: Do not stain the glass surface located deep inside the calibrator.
3 When “OK” is selected in the screen shown Fig. 3, “Calibration Tool's
Value” input screen appears.
4 Input the number shown in the label attached to the calibrator with
using “ ” “ ” “
.
After this input, select “Set” to start calibration.
Calibration is automatically executed in accordance with pre-programmed steps.
Although the number has been already input into the device at the factory, please verify if
the number was input correctly.
Input Calibration Value
Please Press "Set" After Input
2 4 0 . 0
Set
Calibration
Step 2
Fig. 3
Fig. 4
Fig. 1
Calibration
Step 1
Yes No
Open the Cali. Tool the Cover
Calibration START ?
Fig. 2